Operation: Advanced
34
Data Transfer Rate Configuration
To configure the data transfer rate, enter the port configuration menu. Press [1] to
enter the Baud rate setup screen. The display should read "Press Select Mode to tog-
gle baud"and display current baud rate. Press the [Select Mode] key to toggle through
the baud rates (38400, 19200, 9600, 2400, and 1200). Pressing the [Enter] key will
accept the valve and return Inspira to the main screen. Default baud rate is 9600.
Step*
Start, Word, and Stop Bit Configuration
To change the bit configuration, enter the port configuration menu by pressing [2].
The display should read "Define Serial Bits, STRT,WD, STP", and display the current bit
settings respectively. The start bit is fixed at 1 bit and word bits are fixed at 8. Use the
numeric keypad to change the number of stop bits. Press [Enter] to return to the main
screen.
Step*
* Note: See page 33 for steps 1 through 4.