2-19 Calibrating the GA-1121 with the DTP41

Calibrating the GA-1121 with the DTP41
Using the ColorWise Pro Tools Calibrator application and the DTP41
spectrophotometer, you can quickly measure color patches generated by the
copier/printer and automatically download these measurements to the GA-1121.
Before using ColorWise Pro Tools and the DTP41 to calibrate the GA-1121,
N
:
OTE
follow the instructions on
T
GA-1121
O CALIBRATE THE
1. Start Calibrator.
For instructions on starting Calibrator, see
2. Choose X-Rite DTP41 as the measurement method.
3. In the Check Print Settings pane, choose the desired calibration set.
Choose the appropriate calibration set for the type of media you use most often.
For this calibration to take effect, the calibration set must be associated with one
N
:
OTE
or more output profiles. There is no need to make any new associations for the default
calibration set, since it is already associated with the default output profile.
4. In the Generate Measurement Page pane, click Print.
The Print Options dialog box is displayed.
5. Choose the appropriate options, and click Print.
For Page Type, choose either 21 or 34 Sorted Patches.
page A-4
for calibrating the DTP41 spectrophotometer.
DTP41
USING THE
page
2-6.