Table 10 Troubleshooting Table - Olympus SteerROVER User Manual

Steerable motorized scanner
Table of Contents

Advertisement

Problem
Handheld controller
display does not activate
Handheld controller
display is activated, yet
scanner does not drive

Table 10 Troubleshooting table

Possible cause
Input power
requirements not met.
Handheld controller not
plugged into power
controller.
Umbilical cable not
properly connected.
SteerROVER system not
started.
Damaged components in
the handheld controller,
scanner, power
controller, or cabling.
Handheld controller is
not in correct mode for
driving.
Damaged components in
the handheld controller,
scanner, power
controller, or cabling.
10-004209-01EN [Q7750159], Rev. 2, May 2018
Solution
Ensure input power
meets requirements (see
"Power Requirements"
on page 219).
Plug handheld controller
into power controller.
Ensure connectors are
dry, clean and connector
pins are not bent.
Check umbilical cable
connections at both ends.
Ensure connectors are
dry and clean and
connector pins are not
bent.
Start the SteerROVER
system (see "System
Start-Up" on page 57).
Contact Olympus
technical support.
See "Operation" on
page 57 for additional
details.
Contact Olympus
technical support.
Troubleshooting