The ND 261 is designed
for use with HEIDENHAIN
angle encoders
with sinusoidal
output signals. These angle encoders
have one reference
mark or several
distance-
coded
reference
marks.
When a reference
mark is crossed over, it generates
a signal identifying
that position
as a reference
point. After switch-on,
simply crossing
over a reference
mark re-
stores the relationship
between
axis positions
and display values as it was last
defined
by datum setting.
With distance-coded
reference
marks, a maximum
traverse
of only 10" or 20" suf-
fices to restore the datum.
Switch-On
Turn on the power
(switch
located on rear panel).
l
Display shows
-1
l
REF blinks.
~
me...CL
Switch
on reference
mark
evaluation
l
Display shows
the value last assigned
to the
reference
mark position.
l
REF indicator
glows.
5,697
l
Decimal point-blinks.
-
1
Cross over reference
mark
\I
Move the axis until the display becomes
active and the
decimal
point no longer blinks.
If you do
not
wish reference
mark evaluation,
press CL instead of ENT.
Setting the Datum
The datum setting procedure
assigns a display value to a specific axis position
The ND 261 allows you to set two separate
datum points.
Select
datum
1 or 2.
pm1
Enter a value,
such as 40,
I
40
-
J
Confirm
entered
value,
You can switch
from one datum to the other at any time.
Use datum 2 when you want to display incremental
dimensions.