Troubleshooting - FeelElec FY6800 Series User Manual

Fully numerical control dual channel function/arbitrary waveform generator
Table of Contents

Advertisement

Troubleshooting

This chapter lists the commonly encountered failures of FY6800and their
solutions. When you encounter these problems, please solve them following the
corresponding steps below. If the problem remains still, please contact
and provide the device information (Press SYSTEM to get it).
Failure Phenomena
The screen of the
generator is still dark (no
display) after switch on.
CH2 is locked.
Set correctly,
but no waveform output
42
1) Check whether the power is correctly connected.
2) Check whether the power switch has been pulled
in place.
3) Restart the instrument after finishing the above
inspections.
4) If it still does not work correctly, please contact
FeelElec.
1) Check that the signal generator is operating in
synchronous state. Press the SYSTEM -> SYNC
button to enter the synchronization settings
interface to cancel all synchronization parameters.
2) If the problem is still, please restart the generator.
1) Check whether the BNC cable is connected
tightly with CH1 or CH2 connector.
2) Check whether the BNC cable has internal
damage.
3) Check whether the BNC cable is connected
tightly with the test instrument.
4) Check whether the indicators of CH1 or CH2 is
turned on. If not press corresponding button to turn
it on.
5) If the problem is still, please contact FeelElec.
Solutions
FY6800 Series User's Manual
FeelElec