Keithley TEC SourceMeter 2510 Service Manual page 44

Hide thumbs Also See for TEC SourceMeter 2510:
Table of Contents

Advertisement

Models 2510 and 2510-AT
55.
56.
Figure 2-5
I-SS and V-SS calibration
resistor connections
Ω2
Ω4
PREV
DCV
ACV
DCI
ACI
DISPLAY
NEXT
REL
TRIG
STORE RECALL
FILTER MATH
POWER
INFO
LOCAL
CHAN SCAN
CONFIG MENU
Model 2002 DMM
Press ENTER to complete the shorted calibration step, then remove the shorting wires
from the INPUT terminals.
Press EXIT to return to the CAL EXECUTE menu.
SENSE
INPUT
Ω 4 WIRE
HI
INPUT HI
350V
1100V
!
PEAK
PEAK
INPUT LO
2002 MULTIMETER
LO
500V
PEAK
INPUTS
FREQ
TEMP
RANGE
F
R
AUTO
FRONT/REAR
2A 250V
AMPS
RANGE
CAL
EXIT
ENTER
1kΩ Resistor
INPUT F+
INPUT F-
INPUT S+
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
!
CAT
I
MADE IN
OUTPUT
INPUT
U.S.A.
F+ S+ S- F- F+ F- S+ S-
ISOLATION FROM EARTH: 30V MAX.
TRIGGER
LINK
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Calibration
2-17
Connect sense (S)
leads as close
as possible to
INPUT S-
resistor body.
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
ENABLE-DIG I/O
RS-232
LINE FUSE
SLOWBLOW
2.5A, 250V
!
LINE RATING
100-240VAC
50, 60 HZ
90VA MAX
Model 2510

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TEC SourceMeter 2510 and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Tec sourcemeter 2510-at

Table of Contents