3Disc FireCR Flash User Manual page 44

Computed radiography reader
Table of Contents

Advertisement

User Manual
QuantorMed Plus
1.1.7. Step 6: Scan High Dose
The Scan High Dose button is enabled after performing a Scan Mid Dose.
Expose the cassette to the X-ray with the conditions noted below the Scan High Dose
button.
Then, insert the cassette into the reader and click Scan High Dose button to acquire
the fourth calibration image.
The intensity value of the high dose image is displayed beneath the button.
The intensity value of the high dose is displayed in green if the value is within the
acceptable range.
The intensity value of the high dose is displayed in red if the value is not within the
acceptable range, click Accept to continue anyways or Reject to acquire the high dose
image again.
Increase the X-ray dose to increase the mean intensity value or lower the X-ray dose
to decrease the mean intensity value. *If changes are made to technique (Dose)
please note the new values used (KVp, mAs) and input the new data in
C:\QuantorMed+\firecr.ini
Figure 2.21. Scan High Dose
8
TM -721-EN