3Disc FireCR Flash User Manual page 30

Computed radiography reader
Table of Contents

Advertisement

User Manual
FireCR Flash
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
EUT is used exceeds the applicable RF compliance level above, the EUT should be observed to
verifynormal operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the EUT.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V / m.
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the FireCR Flash system.
The FireCR Flash system is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The user of the FireCR Flash system can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the FireCR Flash system as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power (W) of transmitter
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
Immunity test
IEC 60601 Test Level
Conducted RF
3 Vrms, 150 kHz to 80
IEC 61000-4-6
MHz
Radiated RF
3 V/m, 80 MHz to 2.5
IEC 61000-4-3
GHz
Separation distance (m) according to frequency of transmitter
150 kHz to 80 MHz
0.12
0.37
1.17
3.70
11.70
Immunity and Compliance Level
Actual Immunity Level
3 Vrms, 150 kHz to 80
MHz
3 V/m, 80 MHz to 2.5
GHz
30
800 MHz to 2.5
80 MHz to 800 MHz
0.12
0.37
1.17
3.70
11.70
Compliance Level
3 Vrms, 150 kHz to 80
MHz
3 V/m, 80 MHz to 2.5
GHz
TM-401-EN
GHz
0.23
0.74
2.33
7.37
23.30