1.1.5. Step 4: Scan Low Dose
The Scan Low Dose button is enabled after performing a Scan Blank.
Expose the cassette to the X-ray with the conditions noted below the Scan Low Dose
button.
Then, insert the cassette into the reader and click Scan Low Dose to acquire the
second calibration image.
The mean intensity value of the low dose image is displayed beneath the button.
The mean intensity value of the low dose is displayed in green if the value is within the
acceptable range.
The mean intensity value of the low dose is displayed in red if the value is not within
the acceptable range, click Accept to continue anyways or Reject to try again.
Increase the X-ray dose to increase the mean intensity value or lower the X-ray dose
to decrease the mean intensity value. *If changes are made to technique (Dose)
please note the new values used (KVp, mAs) and input the new data in
C:\QuantorMed+\firecr.ini
Figure 2.19. Scan Low Dose