User Manual
QuantorMed Plus
1.1.4. Step 3: Scan Blank
The Scan Blank button is enabled after performing the Erase step.
After the Erase step, insert the cassette with no X-ray exposure into the reader and
click Scan Blank to acquire the first calibration image.
The mean intensity value of the blank image is displayed beneath the button.
The mean intensity value of the blank image is displayed in green if the value is within
the acceptable range.
The mean intensity value of the blank image is displayed in red if the value is not within
the acceptable range.
If the mean intensity value of the blank image is displayed in red, click Accept to
continue anyways or Reject to try again.
To lower the mean intensity value of the blank image, click Erase again.
Figure 2.18. Scan Blank
5