Canon CXDI-50G User Manual page 6

Digital radiography
Hide thumbs Also See for CXDI-50G:
Table of Contents

Advertisement

Safety Information
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflections from structures, object and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can-
not be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength
in the location in which the CXDI-50G is used exceeds the applicable RF compliance level above,
the CXDI-50G should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the CXDI-50G.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
IEC 60601 Test
Compliance
Level
3 Vrms
3V
150 kHz to 80
MHz
3 V/m
3 V/m
80 MHz to 2.5
GHz
Electromagnetic Environment – Guidance
Level
Portable and mobile RF communications
equipment should be used no closer to
any part of the CXDI-50G, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the trans-
mitter.
Recommended separations distance
d
=
1.2 P
d
=
1.2 P 80 MHz to 800 MHz
d
=
2.3 P 800MHz to 2.5 GHz
where P is the maximum output power rat-
ing of the transmitter in watts (W) accord-
ing to the transmitter manufacturer and d
is the recommended separation distance
in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a
survey
, should be less than the compli-
ance level in each frequency range
Interference may occur in the vicinity of
equipment marked with the following sym-
bol:
(4)
b
.

Hide quick links: