Wiring procedure
Perform the following procedure to wire circuitry to
the screw terminal connector card:
1.
2.
3.
4.
5.
6.
7.
WARNING
Make sure all power is off and any
stored energy in external circuitry is
discharged.
Jf mated together, separate the connector card from
the relay card by removing the mounting screw
and pulling the two cards away from each other.
Remember to only handle the cards by the edges
and shields to avoid contamination.
Using an insulated screwdriver, connect the cir-
cuitry to the appropriate terminals. Figure 3-6
shows how the output of Bank A would be con-
nected to a DMM.
Referring to Figure 3-7, remove the top half of the
cable clamp as follows:
A. Loosen the cable clamp screw enough to disen-
gage it from the bottom half of the cable clamp.
B. Using your thumb and forefinger, press the re-
taining clips inward and, with your other
hand, remove the top half of the clamp.
Route wires under wire guide/connector shim.
Route the wires through the bottom half of the ca-
ble clamp.
Replace the top half of the clamp. It simply snaps
onto the bottom half of the clamp. lighten the ca-
ble clamp screw. The clamp serves as a strain relief
for terminal block wires.
Mate the connector card to the relay card. The
Model 7011 is now ready to be installed in the
Model 7001 mainframe. See paragraph 3.5 for de-
tails.
Figure 3-6
Typical screw
terminal
connections
'
Clips \
Figure 3-7
Cable clamp for screw terminal
connector
card
3-5