TEMPERATURE RECORDER

WARNING
Disconnect the power supply plug before attaching the temperature recorder or it may cause
electric shock or fire.
The figure below shows the description of a temperature recorder.
Warning indicator
Door
Feeding of chart
1. Open the door and let down the pen holder lever; the pen point
is apart from the chart. (Fig. 1)
2. Pull the cartridge out of the mounted position. (Fig. 2)
3. Set a new chart in place on the rear bottom of the cartridge.
Set the hole on the chart in the cog of the chart driving assembly
and feed the chart in the direction of the arrow by driving the cog
wheel.
4. Set the chart according to the mark of day and time. (Fig. 3)
5. When mounting, lay down the cartridge first, and push into the
mounting position with the groove set a the projection on the
position.
6. Set up the cartridge and settle in position.
7. Lift up the pen holder lever and close the door.
Replacement of cell
Replace the dry cell once a year as follows:
1. Put down the pen holder lever first, then pull out cartridge from
the mounted position.
2. Reset the battery box cover with wire, on the bottom left side,
and take out the cell.
3. Set a new cell in the case.
4. Shut the battery box cover after exchanging cells. Mount the
cartridge as before and lay up the penholder lever.
Setting of backup temperature
1. There is a warning indicator (red guide) on the top of the recording indicator. Adjust the warning
indicator with your finger to backup temperature, at which temperature the backup cooling kit starts to
operate.
2. Set at a temperature 15
Recording indicator
Pen holder lever
Cartridge
o
C higher than that in the freezer chamber.
27
Time
Fig. 1
Fig. 2
Fig. 3