ALARM TEMPERATURE SETTING

This unit is provided with the high temperature alarm. The setting of high temperature alarm is 10
o
15
C (The setting value is 2 kinds.) higher than the setting of chamber temperature. The procedure in
table 2 shows the sequence to set the high temperature alarm at 15
temperature. Set the figure to 000 at step 5 when setting the high temperature alarm at 10
than the setting of the set temperature.
Note:
The alarm temperature is set at the factory 10
Table 2. High temperature alarm setting (15
Description of operation
1
Press the numerical value shift key
2
for about 5 seconds.
Press numerical value shift key and
3
scroll the figure to 1.
4 Press enter key.
Scroll the figure to 1 by using
5
numerical value shift key.
6 Press enter key.
Note: The set mode returns to the temperature display mode automatically when 90 seconds has
passed without any key operation.
o
C higher than the setting of chamber temperature.
o
C higher than the setting of chamber temperature)
Key operated
-----
Set temperature is displayed.
The first digit blinks.
The first digit blinks.
ENT
The first digit blinks.
When pressed, the figure of settable
digit changes.
Alarm temperature is memorized and
ENT
the current chamber temperature
is displayed.
15
o
C higher than the setting of chamber
Indication after operation
o
C or
o
C higher