Rad-icon 1520 GigE User Manual
cycling will return it to conductive state again thus restoring normal operation (assuming
power has been properly applied).
WARNING: When setting up the detector power supplies follow these guidelines:
!
Note: If your power supply does not meet these requirements, the detector performance
specifications cannot be guaranteed.
Trigger Connector
The trigger circuitry inside the detector is shown in the figure below.
Trigger Output Collector +
Trigger Output Emitter -
Trigger Input Cathode -
1
2
3
4
Apply the appropriate voltages.
Keep leads as short as possible in order to reduce voltage drop. The voltage level
at the product power connector should be within the range mentioned in the
datasheet.
Use high-quality linear supplies of SELV type and/or EN 60601-1 compliant.
1
ESD protection
4 x ESD12VS1UA
2
Trigger Input Anode +
3
4
Shield
Pin 1 indicator
Description
Out
optocoupler collector with 680R +100R
series resistor (note 1)
Out
optocoupler emitter
In
optocoupler anode with 680R series resistor (note 2)
In
optocoupler cathode
Shielding (chassis)
680
100
BAT54J
100p
TCMT1600
680
100p
TCMT1600
11
Teledyne DALSA
Need help?
Do you have a question about the Rad-icon 1520 and is the answer not in the manual?