Installation Procedure for the Upgrade
Option Enable Procedure
1. Locate the email(s) from Keysight which contain license file attachments. These emails are a
result of
"Step 1. Obtain a Keyword and Verify the Information" on page
2. Copy the license file(s) from the email(s) to the root d irectory of the USB flash drive.
More than one license file may be copied to the USB flash drive.
3. Insert the USB flash drive into the PNA's USB drive slot.
4. The instrument automatically reads the license file(s) from the USB flash drive, validates the
file(s), and installs the option(s). Generally, warnings or error messages may be ignored.
5. Wait >1 minute for your Option to be installed.
The instrument will not prompt you when the upgrade has been completed.
6. The PNA application automatically restarts after the option installation has been completed.
Optional: Cycle power on the instrument to activate the newly installed option upgrade(s).
(Press the power button once to power down the instrument.)
Option Verification Procedure
Once the analyzer has restarted and or the Network Analyzer program is again running:
1. On the analyzer's Help menu, press About Network Analyzer.
2. Verify that "422" are listed, but "222" is not listed after "Options:" in the display. Press OK.
Step 33. Perform Post-Upgrade Adjustments and Calibration
Adjustments
The following adjustments must be made due to the hardware changes of the analyzer.
— source adjustment
— IF gain adjustment
39
A single license file may contain licenses for more than one instrument.
Attempting to re-install a license file that is already installed may generate
a "Corrupt Media" error message. Ignore this message.
If the option(s) have not been enabled, contact Keysight Technologies.
Refer to
"Getting Assistance from Keysight" on page
13.
5.
Installation Note N5242-90123