Download Print this page

Keysight N5241BU-622 Installation Note page 27

Add 4-port capability upgrade kit to upgrade pna-x n5241b, n5242b or n5249b option 222 to option 422

Advertisement

Installation Procedure for the Upgrade
Step 20.Install the Test Set Cables
Flexible Cables Required for Upgrading to an Option 422 PNA With or Without Option 029
Install the following flexible cables in the order listed. To see images showing the location of these
cables, click either of the Chapter 6 bookmarks "Bottom RF Cables, 4-Port, Option 422..." in the
1
PDF Service Guide
. New parts are listed in
— W56 (N5242-60019) A24 mixer brick (R4) to A20 IF multiplexer (P414)
— W57 (N5242-60020) A24 mixer brick (R3) to A20 IF multiplexer (P413)
— W52 (N5242-60021) A23 mixer brick (R1) to A20 IF multiplexer (P411)
— W53 (N5242-60022) A23 mixer brick (R2) to A20 IF multiplexer (P412)
— W58 (N5242-60023) A24 mixer brick (C) to A20 IF multiplexer (P601)
— W55 (N5242-60024) A24 mixer brick (D) to A20 IF multiplexer (P801)
Semirigid Cables Required for Upgrading to an Option 422 PNA With Option 029
If Option 029 is not installed, proceed to
Install the following semirigid cables in the order listed if upgrading to a PNA Option 422 with
Option 029. To see images showing the location of these cables, click the Chapter 6 bookmark
"Bottom RF Cables, 4-Port, Option 422 with Option 029" in the PDF Service Guide
listed in
Table 1 on page
— W26 (N5222-20034) A32 port 2 coupler to front-panel Port 2 CPLR ARM
— W135 (reuse) (N5242-20073 or N5242-20278) Front-panel Port 2 RCVR B IN to A45 port 2
receiver attenuator
1. See
"Downloading the Online PNA Service Guide" on page
27
Follow instructions carefully when making cable connections, especially
wire harness connections. Incorrect connections can destroy components,
resulting in additional customer costs.
Be careful not to damage the center pins of the semirigid cables. Some
flexing of the cables may be necessary but do not over-bend them.
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections
except the front and rear panel bulkhead connectors. On these, use a 9
mm nutsetter or open end torque wrench set to 21 in-lb.
Cables that are to be reinstalled are designated with "reuse." Part numbers
in a normal weight font are for analyzers with serial numbers
MY/SG/US5310 and below. Part numbers in a bold weight font are for
analyzers with serial numbers MY/SG/US5321 and above.
page
9.
Table 1 on page
9.
31.
7.
1
. New parts are
Installation Note N5242-90123

Advertisement

loading

This manual is also suitable for:

N5242bu-622N5249bu-622