3 GETTING STARTED (continued)
The Elcometer 130 Model S does not have automatic temperature compensation and therefore, the
actual level of soluble salts needs to be calculated manually using the following equation:
k =
Example: Measured Value: 12.3
k =
If using the Elcometer 130 Model T with Automatic Temperature Compensation disabled, the same
equation should be used to correct the gauge reading for temperature variances.
7
k
measured
1 + c [T - T ]
cal
2
µg/cm
at 28°C
12.3
1 + 0.02 [28 - 25]
Where:
k = Corrected Result
c = Temperature Coefficient (2.0%)
T = Measurement Temperature °C
T = Calibration Temperature (25°C)
cal
2
11.6 µg/cm