Elcometer 130 User Manual page 10

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
R
4 MEASURING SOLUBLE SALTS (continued)
8
After 2 minutes, remove the filter paper from the test surface and place on
to the measurement electrodes.
4
For half size paper: position on the bottom half of the circle using the markers on
the measurement electrodes as a guide.
4
For quarter size paper: position on the bottom right quadrant of the circle using the
markers on the measurement electrodes as a guide.
9
Close the lid ensuring that the magnetic catch is fully engaged.
10
The reading will automatically be displayed on screen and stored into
memory together with the filter paper size, temperature, date and time
(Model T - when in Batch Mode).
4
The Elcometer 130 will automatically adjust the reading according to the filter
paper size detected or selected, assuming that the paper is positioned on the
measurement electrodes correctly, and if manually set up, the correct paper size is
used. No manual calculations are required by the user.
11
Lift the lid and remove the filter paper.
12
Place the filter paper in a resealable bag (supplied), if required for further
chemical analysis.
13
Clean the electrodes between tests using pure water and a sensor wipe
(supplied).
Failure to thoroughly clean the electrodes between tests may contaminate
subsequent tests and result in inaccurate readings.
The Elcometer 130 measurement electrodes are gold plated to prevent corrosion
and oxidisation, prolonging the life and accuracy of the gauge. They should be
cleaned using pure water and the sensor wipes supplied. DO NOT use abrasive
materials as this will damage the electrodes.
9
a
b
c
www.elcometer.com