Fundus Photography – Colour or Auto Fluorescence
Position the Patient - Set the Patient's chin on chinrest. Keep their chin and forehead stable.
Be sure to adjust the chinrest height to align the eye marker and the corner of the eye.
IR Illumination
If IR fundus image
is noisy or dark,
please turn up the
illumination level.
1. Select
[Colour]
[Auto Fluo]
depending on
the Fundus Photo required.
6
Once aligned and evenly illuminated, instruct the Patient not to blink and
Instruct the Patient to look straight ahead and find the internal
press the trigger on the joystick to capture.
fixation target.
7
Press the
Select
[Fundus Photo]
Press the relevant
[Fundus Photo]
change fixation position from Macula, Centre or Disc.
Move the machine in until the two circles are green and together within the brackets.
Adjust IR brightness or flash power if required and also the fixation target to adjust
Patient fixation.
12
opcon (Great Britain) Medical Limited
12
Topcon (Great Britain) Medical Limited
or
button on the PC to save your images to Imagenet 6.
[Save]
on main menu.
button –
[Colour]
| DRI OCT Triton | Quick Reference Guide |
Adjust the Patient's height using the
and
[Down]
chinrest buttons found next to
the joystick on the machine.
2. Change fixation position
if required from Macula,
Centre or Disc.
or
[Auto Fluorescence]
| DRI OCT Triton | Quick Reference Guide |
Fundus Photography
3. If the Patient's fixation
needs adjusting, this little
blue cross is the green
internal fixation cross.
4. Ensure distance
indicator circles are green
and overlapping and in
the brackets.
Yellow = too close
Orange = too far away
[Up]
5. Flash Intensity
Increase if the Patient
is poorly dilated or
has small pupils.
-
Fundus Photography