System Information Used In Troubleshooting - DENTSPLY Multimat2Sinter Instructions For Use Manual

Sintering furnace
Table of Contents

Advertisement

Problem
Display shows "Door
open" even though
furnace door is closed
(lifting plate is up)
9.4

System information used in troubleshooting

To be performed by operator
Possible cause
RCD too sensitive.
Terminal position switch
stuck or broken.
Controller hardware or software problems require assistance by
our service technician ( Table of troubleshooting messages,
page 39).
Please provide the following system information to the service
technician for troubleshooting:
Serial number
Hardware version
Software version
View system information as follows:
1.
Turn on the furnace with the On/Off switch.
Values and/or texts will appear on the display for about 3
seconds.
2.
Record this information and notify the service technician.
Page 40
Potential remedies
Install 30-mA RCDs.
Visual inspection,
Contact service technician.
Cleared by
Service
technician