3 Technical specifications
3.1
General
Parameter
External dimensions (W × H × D)
Heating chamber volume
Maximum temperature
Heating rate
Weight
3.2
Electrical connection
Electrical
3.3
Operating conditions
Environment
Value
385 × 780 × 500 mm
2 bowls 120 mm × 35 mm
1650 °C
selectable: 1–99 °C/min
Parameter
Nominal supply voltage
Frequency
Permissible deviation
from the nominal voltage
Maximum Power
consumption
Overvoltage category
Protection against
overvoltage
Fuse rating (on site)
Parameter
Temperature range
Maximum relative
humidity
Installation
Height above sea level
Pollution severity
Page 11
Unit
58 kg
Value
220-240
± 10
2000
II (pursuant to IEC 60364-4-443)
Provide suitable surge filters
Connect to a dedicated circuit protected by a
T10 AH 250 V fuse
Value
5–40
up to 31°C: 80
up to 40°C: 50
non-condensing
Indoors in dry rooms only
maximum 2000
Tested for pollution severity – Pollution degree II
Unit
V AC
50
Hz
%
W
Unit
°C
%
%
m