Hach 5500sc Operations page 15

Ammonia/monochloramine
Table of Contents

Advertisement

Available languages

Available languages

Option
OVERFEED TIMER Sets the maximum amount of time to get to the process setpoint. When the time
OFF DELAY
ON DELAY
5. Select EVENT CONTROL in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
Description
SET SETPOINT Sets the value to set the relay to on.
DEADBAND
Sets a delay so that the relay will be stable as the process value converges to the setpoint.
OnMax TIMER
Sets the maximum time the relay stays on (default = 0 min).
OffMax TIMER
Sets the maximum time the relay stays off (default = 0 min).
OnMin TIMER
Sets the time the relay stays on, independent from the measured value (default = 0 min).
OffMin TIMER
Sets the time the relay stays off, independent from the measured value (default = 0 min).
6. Select SCHEDULER in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
HOLD OUTPUTS Holds or transfers outputs for the selected channels.
RUN DAYS
START TIME
INTERVAL
DURATION
OFF DELAY
7. Select WARNING in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
WARNING LEVEL
8. Select PROCESS EVENT in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
MEASURING 1
MEASURING 2
CLEANING
ZERO CAL
SLOPE CAL
SHUTDOWN
STARTUP
GRAB SAMPLE
Description
expires and the relay does not show the setpoint, the relay is set to off. After an
overfeed alarm occurs, reset the timer manually.
Sets a delay time to set the relay to off (default = 5 seconds).
Sets a delay time to set the relay to on (default = 5 seconds).
Description
Selects the days the relay stays on. Options: Sunday, Monday, Tuesday, Wednesday,
Thursday, Friday, Saturday
Sets the start time.
Sets the time between activation cycles (default = 5 min).
Sets the period of time the relay is set to on (default = 30 sec).
Sets the time for additional hold/output time after the relay is set to off.
Description
Sets the level for the warning activation and starts applicable individual warnings.
Description
Sets the relay to on when cell 1 is measuring.
Sets the relay to on when cell 2 is measuring.
Sets the relay to on when a cleaning cycle is started.
Sets the relay to on when a zero calibration has started.
Sets the relay to on when a slope calibration has started.
Sets the relay to on when a shutdown cycle is started.
Sets the relay to on when a startup cycle is started.
Sets the relay to on when a grab sample cycle is started.
English 15