Etude Plus
Dismounting
1.
Lower the side rail.
2.
Dismount one end at a time. Lift the lower bar, so that the spring
pawl is visible.
3.
Press the spring pawl with a key or a screwdriver.
4.
Hold the lower bar. Lower the side rail, until all three bars are
free of the bed end rail.
5.
Repeat points 2 to 4 when dismounting the other end of the side
rail.
Operating
As a standard, the side rails can only be locked in top position.
1.
To lock / raise to upper position: Pull up the top side rail bar,
until the locking pin locks with an audible click
2.
To release to lower position: Lift the top side rail bar and
press the two locking rings together and lower the side rail.

5.3 Lifting pole

WARNING!
Risk of injury
If the lifting pole is used while the handle has been turned
away from the bed, the bed can tip when the handle is
used.
– Position the lifting pole in such a way that the handle
extends inwards across the bed.
Mounting
1.
Remove the plug from the lifting pole tube at the head end of the
bed where the lifting pole is to be placed.
2.
Insert the lifting pole into the lifting pole tube and fix with the
thumb screw.
Adjusting the height of the lifting pole handle
10
1.
Loosen the cord.
2.
The lifting handle can now be adjusted to the desired height.
3.
Press the cord together and check that the cord is locked in the
cord lock by pulling the handle.
The distance between the lifting pole and the mattress
Minimum (at 16 cm mattress)
Maximum (at 10 cm mattress)

5.4 Mattress support extension

For patients, taller than two meters it is recommended to
use a mattress support extension.
Fitting the mattress support extension
1.
Disassemble the bed and remove the inserts.
2.
Mount the mattress support extension between both parts of the
mattress support and reassemble the bed.

5.5 Mattress

WARNING!
Safety aspects regarding combination of side rails
and mattresses:
– In order to obtain the highest possible safety level,
when using the bed in combination with side rails, it
is important to respect the minimum and maximum
measures for mattresses stated in the table.
– To mitigate the possibility of the user crawling over the
top of the side rail, with the risk of severe injury due
to e.g. falling, a minimum vertical distance of 22 cm
between the top of the mattress and the top of the
side rail are to be respected. The maximum mattress
height in combination with a given side rail is stated
in the table.
– To prevent head entrapment between the inside
surface of the side rail and the mattress, with the risk
of e.g. suffocation, a safety maximum space has to be
respected. When especially very thick mattresses or
soft mattresses (low density) or a combination of these
it is important to focus on this risk. The minimum
width (and length) of mattresses in combination with a
given side rail is stated in the table.
Side rail:
height
Scala Basic 2
Scala Basic Plus 2
Scala Medium 2
Scala Decubi 2
Line*, mattress support
platform in low position
Line*, mattress support
platform in high position
Britt V, mattress support
platform in low position
Britt V, mattress support
platform in high position
36-59 cm
42-65 cm
Mattress measures (in cm):
Min.
Max.
Min.
height
width
10
17
85 – 90
10
17
85 – 90
10
22
85 – 90
23
31
85 – 90
11
22
88 – 90
5
15
88 – 90
11
22
85 – 90
5
15
85 – 90
Min.
length
200
200
200
200
200
200
200
200
1567454-A