Karl Storz UP 501 S-PILOT Instruction Manual page 52

Tube squeeze valve
Table of Contents

Advertisement

44
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Cables which have been shown to comply with IEC 60601-1-2:
Cable type
Shielded
PA
No
SCB Cable
Yes
20 0141 30
Yes
27 6100 70
Yes
20 7010 70
Yes
UP 004
No
UP 005
No
Power Cord
No
Guidance and manufacturer's declaration – electromagnetic emissions
The S-PILOT™ model UP 501 is intended for use in the electromagnetic environment speci ed below.
The customer or user of the S-PILOT™ model UP 501 should ensure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Electromagnetic
Compatibility (EMC)
information
Length [m]
Ferrite
> 3
No
> 1
No
3
No
0.42
No
1
No
3
No
3
No
3
No
Table 1
Compliance
Group 1
The S-PILOT™ model UP 501 uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause interference in nearby
electronic equipment.
Class B
The S-PILOT™ model UP 501 is suitable for use in all establishments including
domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Class A
Complies
Indicaciones sobre
compatibilidad
electromagnética (CEM)
Used for
Potential equalization
Connecting SCB network, other SCB units
One-pedal-Footswitch
Control cable (connecting cable between the S-PILOT™ and
the UNIDRIVE
®
S III)
Control cable (connecting cable between the S-PILOT™ and
UNIDRIVE
®
S III /CALCUSON)
Connecting cable between the S-PILOT™ and
AUTOCON
®
III 400/ConMed system 2450 or 5000
Connecting cable between the S-PILOT™ and Covidien
Force Triad or Covidien Valleylab Force FX
Power supply
Electromagnetic environment – guidance