Device Command List For The Edge-Displacement Signal - HP 81110A Quick Start Manual

Pulse/pattern generators
Table of Contents

Advertisement

Device Command List for the Edge-displacement
Signal
NOTE:
# Reset the instrument to start from a defined, default status.
*RST
# Switch off the automatic display update to increase programming
# speed.
:DISPlay OFF
# Set the instrument to pattern mode and select NRZ data output format
# for output 1 and output 2.
:DIGital:PATTern ON
:DIGital:SIGNal1:FORMat NRZ
:DIGital:SIGNal2:FORMat NRZ
# Define a pattern length of 8 bit and program for channel 1 the pattern
# 10100010, for the channel 2 the pattern 00001000, see manual step 3.
:TRIGger:COUNt 8
:DIGital:PATTern:DATA1 #1810100010
HP 81110/04A Quick Start
If you want to include this Edge-displacement Signal in your pro-
gram use the following command lines in boldface font. The com-
ment lines starting with a # are not required.
Two Outputs are required for this example. As channel addition
feature is used this type of signal can be performed by HP 81104A
and HP 81110A with HP 81111A 10V/165 MHz Outputs.
Getting Started
Getting Started
Examples
95

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

81104a

Table of Contents