Troubleshooting - Chicco Easy Touch Instructions For Use Manual

Infrared forehead thermometer
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 14
Error Message

Troubleshooting:

Problem
Device stabilization in process. Wait until head stops flashing.
Battery is low and no more me-
asurements are possible.
Measurement before device
stabilization.
The device showing a rapid am-
bient temperature change.
The ambient temperature is not
within the range between 10°C
and 40°C (50°F~104°F).
Error 5~9, the system is not
functioning properly.
Temperature taken is higher
than 42.2°C (108.0°F) for fore-
head mode or 80°C (176°F) for
scan mode.
Temperature taken is lower
than 34°C (93.2°F) for forehe-
ad mode or –22°C (-7.6°F) for
scan mode.
Device can not be powered on
to the ready stage.
Replace the battery.
Wait until head stops flashing.
Allow the thermometer to rest in a
room for at least 30 minutes at room
temperature:
10°C and 40°C(50°F~104°F).
Allow the thermometer to rest in a
room at least 30 minutes at room tem-
perature: 10°C and 40°C(50°F~
104°F).
Unload the battery, wait for 1 minute
and repower it. If the message reappe-
ars, contact the retailer for service.
1) For forehead mode, please stay in
a stable environment for 5mins
and avoid the exercise and bath for
30min, then make sure the probe lens
is clean before take a new temperatu-
re measurement.
2) For scan mode, please select target
within specifications. If a mal-fun-
ction still exists, please contact the
nearest retailer.
Change with a new battery.
20
Solution

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents