Servicing Sirona Quick Coupling - Dentsply Sirona Midwest Stylus Plus Operating Instructions Manual

Air-driven-high-speed handpiece/turbine
Hide thumbs Also See for Midwest Stylus Plus:
Table of Contents

Advertisement

Available languages

Available languages

6607290.book Seite 54 Montag, 5. September 2016 11:53 11
8 Maintenance
A
B
54
Operating Instructions Midwest Stylus
8.3

Servicing Sirona quick coupling

8.3.1
Replacing O-rings
1. Remove the defective O-ring.
Using T1 Spray
O-Ring Removal/Attachment _ Using Tool
2. Slide the tool (A), with the new O-ring attached, to just in front of the
corresponding groove (B).
3. Insert the O-ring. Do not use any sharp objects.
4. Lightly oil the O-ring with Midwest
NOTICE! Do not use Vaseline or silicone grease on the O-rings.
®
Plus Air-Driven-High-Speed Handpiece
®
Plus Aerosol Spray.
6607290 D3673.201.01.01.02

Advertisement

Table of Contents
loading

Table of Contents