Konica Minolta SRX-101A Service Manual page 60

Medical film processor
Hide thumbs Also See for SRX-101A:
Table of Contents

Advertisement

6-2-2. DIP Switch Settings
1. DS1
SW
Function
1
Mode
2
Power source frequency
3
Selection of
4
processing time
5
Selection of
6
standby mode
7
8
Selection of
replenisher correction Rate of replenisher
2. DS2
SW
position
[0]
[1]
[2]
[3]
[4]
[5]
[6]
[7]
[8]
[9]
[A]
[B]
[C]
[D]
[E]
[F]
Normal operation mode
DS1-3
OFF
DS1-4
OFF
Processing Time
90 sec.
DS1-5
OFF
DS1-6
OFF
Standby Time
0 min.
DS1-7
OFF
DS1-8
OFF
0%
correction
Chemical Mixer
Dryer Pre-Heat Temp. (:)
Available
Not available
CPU BOARD DIP SWITCHES
OFF
Maintenance Mode
50 Hz
60 Hz
ON
OFF
OFF
ON
120 sec.
180 sec.
ON
OFF
OFF
ON
10 min.
30 min.
ON
OFF
OFF
ON
-20%
-40%
Water Replenisher(cc)
DEV
0
40
230
75
100
0
40
250
75
100
0
40
230
75
100
0
40
250
75
100
ON
ON
ON
Use prohibited
ON
ON
Use prohibited
ON
ON
30%
FIX
0
20
35
45
0
20
35
45
0
20
35
45
0
20
35
45
SERVICE MANUAL
55