Time Chart - Konica Minolta SRX-101A Service Manual

Medical film processor
Hide thumbs Also See for SRX-101A:
Table of Contents

Advertisement

4. Time Chart (In normal functioning)
"RUN" Lamp
"READY" Lamp
READY Signal
Drive Motor
DEV Circulation
Pump
FIX Circulation
Pump
53.3s
DEV Supply Pump
(44.4s)
46.7s
FIX Supply Pump
(38.9s)
DEV Heater
13min
DRY Heater
Preheating
Alarm
2min
Electromagnetic
Valve
8min
(
)
Pump to refill water
Drying Fan
Input
Notes :
1) T1 is one of the three standby options : 10 minutes, 30 minutes, or 0minute (continuous operation).
Ex. : When a film is inserted while a 10 minutes' standby interval, a 10 minutes' timer is reset.
Then after the film is ejected, the T1 repeats 5 minutes ON and 10 minutes OFF alternately.
2) The various for T2 (duration of high speed rotation of the motor after the film detection is turned off)
and T3 (duration from detection of the trailing edge of film to alarm ON) are listed below :
Cycle Time
90
120
180
3) The time in the chart is valid at the 50 Hz. ( ) is at the 60 Hz.
4) For each hour after [RUN] SW ON during which no film is processed, one 24 x 30cm(10 x 12inch)
replenish amount of processing solution will be supplied to each tank. (The timer is reset after Film
Insertion Detection ON.)
5) Drying heater preheating will begin if the temperature is below 80: at [RUN] SW ON.
6) Initial solution replenish will not be carried out if the [RUN] SW is pressed ON within 4 hours after
[RUN] SW OFF.
7) Initial solution replenish (developer ; 26.7 sec(22.2 sec ; 60 Hz)(two 24 x 30cm(10 x 12inch) Sheets
equivalent) ; fixer solution : 23.3 sec (19.4 sec : 60 Hz)[one 24 x 30cm(10 x 12inch) Sheet equivalent]
will be carried out if the [RUN] SW is pressed ON more than 4 hours after [RUN] SW OFF.
8) Even if DEV and DRY is ready, "READY" lamp doesn't lit at least 15 minutes.
9) Shutdown will occur automatically if film isn't inserted for more 8 hours after "READY" lamp ON.
15min
1h
13.3s
(11.1s)
23.3s
(19.4s)
T1
5
min
5
min
24s
(20s)
60s
(50s)
T2
T3
110
14.0
150
6.3
220
5.0
(Rapid Transfer)
30min
13.3s
3min
(11.1s)
23.3s
(19.4s)
1.5s
T2
60s
T3
(unit : second)
NAME OF PARTS
2min
(Optional)
SERVICE MANUAL
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents