3) Solution Temperature Control Tank
(1) Components Diagram
From the Fixer
Circulation Tank
FIX Tank
From the Developer
Circulation Pump
(2) Capabilities
1) Developer temperature control is carried out by warming with a heater and by natural
cooling.
2) Fixer temperature is maintained by means of heat transferred from the developer
temperature control tank.
3) Circulation volumes :
At 50 Hz
At 60 Hz
4) Rates of Temperature Rise :
Developer : approx. 1.07:/minute
Fixer
5) Sensor Precision :
Developer : +/-0.5: at settings between 32: and 34:.
Heater
: Developer 4.0 liters ; Fixer 4.0 liters/minute
: Developer 4.8 liters ; Fixer 4.8 liters/minute
: approx. 0.75:/minute
NAME OF PARTS
Temperature Control Tank
To the DEV Tank
Thermometer
SERVICE MANUAL