No.
Trouble
6
Discoloration
· The fixer solution temperature
is too low.
· The wash water temperature
is too low.
· The fixer is exhausted due to
overuse. (when processing
small film lots.)
· There is no water in the
WASH Water tank.
7
Insufficient drying · Incorrect temperature setting.
The drying heater (H1) is
faulty.
The drying heater fans (FM1,2)
are faulty.
The drying sensor (RB1)
is faulty.
8
Uneven drying
· Incorrect temperature setting.
The drying heater (H1) is
faulty.
The drying heater fans (FM1,2)
are faulty.
The drying sensor (RB1) is
faulty.
· The wipe-down roller is dirty.
Cause
Response
· Replace the fixer
solution.
· Replace the faulty
parts.
· Correct the temperature
setting.
· Replace the faulty parts.
· Correct the temperature
setting.
· Replace the faulty parts.
· Clean or replace
the wipe-down roller.
TROUBLE SHOOTING
Advice To Customer
· Show the proper method for
maintaining optimum fixer temperature.
· Install a water heating device.
· Remind about processing solution
replacement intervals.
· Show checking method of the water
supply valve and wash water drainage
valve.