Konica Minolta SRX-101A Service Manual page 45

Medical film processor
Hide thumbs Also See for SRX-101A:
Table of Contents

Advertisement

TROUBLE SHOOTING
17. The READY lamp will
not come ON.
Has 15 minutes passed since
the "RUN" lamp came ON?
YES
Has initial wash water
supply been completed
(LED5 OFF)?
YES
Is the film entry detector ON
(The film insertion alarm will
sound).
NO
Is 0.5V or less being supplied to
JJ2 #4 and #7?
YES
Replace the Control Panel
board.
18. The READY lamp will
not turn OFF (RUN lamp)
Has 1 minute passed since
all films being processed
and exitted?
YES
Replace the CPU board.
40
SERVICE MANUAL
NO
Wait the required time.
NO
Wait the required time.
(about 2 minutes)
Check if there is any film
YES
jammed at the detector unit
or not.
NO
Replace the CPU board.
NO
Wait at least 1 minute, and
then press the [RUN] SW.
If all of the checks are OK,
replace the film entry
detector unit.

Hide quick links: