INSTALLATION PROCEDURE
7) Cut a piece of Packing(D)[0thickness 10mm] into a strip of 530mm long. Attach it to the Lower Light
Shield as shown in the diagram below. Attach 4 clip nuts at the top edge of the Lower light Shield,
adjusting the position against the hole, fix the Lower Light Shield to the main unit stand with M6 x
20 pan-head screws, flat washers(φ6) and hexagonal nuts.
Packing
"A" Enlargement
8) Cut two pieces of Packing(A)[thickness 3mm] into strips of 165mm long, and attach them to the
bottoms of the Left and Right Shield. Cut two more pieces of Packing(A)[thickness 3mm] into strips
of 225mm long and attach them to the sides of the Left and Right Light Shields as shown in the
diagram below. Attach two clip nuts onto each top edge of the Left and Right Light Shields adjusting
the position against the hole. Using four M4 x 6 pan-head screws from the light room side of the
shields to fix the Left and Right light Shields to the Lower Light Shield.
Note : The right Light Shield is the one with the manufacturing date stamped on it.
Speed Nuts
M4x6 Pan Screw
Left Light Shield
166mm Strip of Packing
137
SERVICE MANUAL
Attach the packing here
Holes for attaching the shield to the stand
225mm Strip of Packing (A)
Lower Light Shield
Attach clip nuts
Attach clip nuts
Lower Light Shield
165mm Strip of Packing (A)
Speed Nuts
Right Light Shield
M4x6 Pan Screw