Specifications; Loop Power - Fluke 715 Instruction Sheet

Volt/ma calibrator
Hide thumbs Also See for 715:
Table of Contents

Advertisement

Specifications

Specifications are based on a one year calibration cycle and
apply from +18°C to +28°C unless stated otherwise. "Counts"
means number of increments or decrements of the least
significant digit.
DC V Input and Output
Range
100 mV
10 V
Input impedance: 2 M Ω (nominal), < 100 pF
Overvoltage protection: 30 V
Voltage drive capability: 1 mA
DC mA Input
Range
24 mA
Overload protection: 125 mA, 250 V fast acting fuse
DC mA Output
Range: 0 mA to 24 mA
Percent display: 0% = 4 mA, 100% = 20 mA
Accuracy: ±(0.015% of reading + 2 counts)
Source mode:
Compliance: 1000 Ω at 20 mA for battery voltage ≥6.8 V (700 Ω
at 20 mA for battery voltage 5.8 to 6.8 V)
Simulate mode:
External loop voltage requirement: 24 V nominal, 30 V maximum,
12 V minimum

Loop Power

24 V ±10%
Resolution
0.01 mV
0.001 V
Resolution
0.001 mA
Accuracy,
±(% of Reading + Counts)
0.02% + 2
0.02% + 2
Accuracy,
±(% of Reading + Counts)
0.015% + 2