Manufacturer's Declaration Table

General Notes

136
The information contained in this section (such as separation distances) is in general specifically written with
regard to the Camino ICP Monitor. The numbers provided will not guarantee faultless operation but should
provide reasonable assurance of such.
Medical electrical equipment requires special precautions regarding electromagnetic compatibility (EMC) and
needs to be installed and put into service according to the EMC information provided in this document and the
remainder of the instructions for use with this device. Failure to follow the precautions regarding EMC may result
in temporary loss of measurement accuracy or disabling of the monitor.
Portable and mobile RF communications equipment can affect medical electrical equipment.
Cables and accessories not specified within the instructions for use are not authorized. Using other cables and/or
accessories may adversely impact safety, performance and electromagnetic compatibility (increased emission
and decreased immunity).
Warning
Care should be taken if the equipment is used adjacent to or stacked with other equipment; if adjacent or stacked use
is inevitable, the equipment should be observed to verify normal operation in the configuration in which it will be used.
Camino ICP Monitor uses a mains plug as a means of isolation to the mains power supply. Do not position the
monitor in a place difficult to operate the mains plug.
While using a defibrillator along with the Camino ICP Monitor, follow the guidance provided by the defibrillator
manufacturer.