Chapter 4 Repair procedures
If you are replacing the cable (part #18):
3
4.5
The subplate components must be removed (Section 4.4) before replacing the subplate.
Obtain the replacement and reverse all steps in Section 4.4 to reinstall major components.
1
Reverse all steps in Section 4.3 until the UUR is completely assembled.
2
Return to Chapter 2 and perform the safety tests and performance verification and
3
adjustment.
4.6
Before removing components from the enclosure, disassemble the UUR by following the pro-
cedures in Sections 4.1 through 4.3. The following sections discuss removal of the enclosure
components, which are shown in Figure 4-5 on page 39.
Removing the enclosure fan (Section 4.6.1)
Removing the AC power entry module (Section 4.6.2)
Removing the rocker switches (Section 4.6.3)
TABLE 4-2
Tools required
Screwdriver
Hex driver
Hex driver
38
neoBLUE® LED Phototherapy System Service Manual
PN 051296
Obtain the replacement and reverse the steps in this section ( 4.4) to install.
Reverse all steps in Section 4.3 until the UUR is completely assembled.
Return to Chapter 2 and perform the safety tests and performance verification and
adjustment.