Setting 0 % And 100 % Output Parameters; Stepping And Ramping The Output; Manually Stepping The Ma Output - Fluke 725Ex User Manual

Multifunction process calibrator
Hide thumbs Also See for 725Ex:
Table of Contents

Advertisement

Setting 0 % and 100 % Output
Parameters
For current output, the Calibrator assumes that 0 %
corresponds to 4 mA and 100 % corresponds to 20 mA.
For other output parameters, the 0 % and 100 % points
must be set before use of the step and ramp functions.
Proceed as follows:
1.
If necessary, press
2.
Select the desired source function and use the arrow
keys to enter the value. Our example is temperature
source using 100 °C and 300 °C values for source.
Enter 100 °C and press and hold
3.
value.
Enter in 300 °C and press and hold
4.
value.
info@Fluke-Direct.com
for SOURCE mode.
to store the
to store the
Fluke-Direct
Mulitfunction Process Calibrator

Setting 0 % and 100 % Output Parameters

This setting may now be used for the following:
Manually stepping an output with 25 % increments.
Jump between the 0 % and 100 % span points by
momentarily pushing

Stepping and Ramping the Output

Two features are available for adjusting the value of
source functions.
Stepping the output manually with the
keys, or in automatic mode.
Ramping the output.
Stepping and ramping apply to all functions except
pressure, which requires use of an external pressure
source.

Manually Stepping the mA Output

To manually step current output, do the following:
Use
or
to step the current up or down in
25 % steps.
Touch momentarily either
go to 100 %.
.com
or
.
and
to go to 0 %, or
to
39
1.888.610.7664

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents