Calibrating a Transmitter

Use the measurement (upper display) and source (lower
display) modes to calibrate a transmitter. This section
applies to all but pressure transmitters. The following
example shows how to calibrate a temperature transmitter.
Connect the Calibrator to the instrument under test as
shown in Figure 19. Proceed as follows to calibrate a
transmitter
1.
Press
press
2.
Press
pressing this key to select the desired thermocouple
type.
3.
If necessary, press
info@Fluke-Direct.com
for current (upper display). If required,
again to activate loop power.
(lower display). If desired, continue
for SOURCE mode.
Fluke-Direct
Mulitfunction Process Calibrator
4.
Set the zero and span parameters by pressing
keys. Enter these parameters by pressing and
holding
and
. For more information on
setting parameters, see "Setting 0 % and 100 %"
earlier in this manual.
5.
Press
or
to select the appropriate location
6.
Perform test checks at 0-25-50-75-100 % points by
pressing
or
. Adjust the transmitter as
necessary.