Preparation For Installation; Wall Mounting - lopital Amfora Instructions For Use Manual

Shower stretcher & changing table
Table of Contents

Advertisement

Installation
Determining the working height
The shower stretcher/changing table allows carers to adjust it to the optimum ergonomic
working height with an adjustment range of 60 cm. The minimum height of the surface of
the stretcher is 40 cm from the floor. The maximum height of the surface depends on the
height of the bathroom.
Maximum height stretcher surface = Height bathroom (minimum 230 cm) – 130 cm. Draw
the hole pattern on the wall in accordance with the attached installation drawing.
We recommend drawing the bottom two holes at a height of 77 cm above the tiled floor,
which creates a working height range for the surface of the stretcher from 40 - 50 cm to
100 - 110 cm from the tiled floor.

Preparation for installation

• The wall where the shower stretcher/changing table is to be mounted must be at least
2.3 metres wide and 2.3 metres high and free of obstacles.
• Draw a plumb line on the bathroom wall where the middle of the shower stretcher/
changing table must be. Keep a minimum of 1.15 metres on both sides of this line free
of obstacles.
• Drill the holes at the locations marked on the wall.

Wall mounting

• Fasten the wall frame with the fasteners that are appropriate for this wall. Ensure that
the tubes of the wall frame are level.
• If you have ordered the 61002400/01 first fit the fixed side panel to the black panels of
the wall frame.
• Fit the shower stretcher/changing table to the wall frame. First fasten the gas springs
and then the hinge points.
• Fit the cover cap last and check that the cap does not touch the wall, ceiling, floor or
skirting board, and that cables are not trapped. The stretcher/changing table must not
be obstructed by any obstacle.
Shower stretcher & changing table | Instructions for use | 9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vienna

Table of Contents