Available languages

Part no. 078G6100
Read these instructions carefully and completely before using
for the fi rst time. Keep these instructions in a safe place.
1

Product description

Hinged cleaning aid for cleaning the eFlow
nel opening and receptacle for fi lling 5 ml isotonic sodium chloride solution
(0.9%). The nebuliser connection cord and the control unit of the eFlow
appliance are also required for operation. Information on connecting the
control unit to the power supply and on operation can be found in the Instruc-
tions for Use of the eFlow
2

Intended use

The easycare cleaning aid can contribute to a shortening
of the nebulisation time if the performance of the eFlow
aerosol head is deteriorating.
As a result of easycare cleaning, the pores in the membrane
of the eFlow
aerosol head are mechanically rinsed free by
®
means of backwashing in the opposite direction.
3
Important information for use
Cleaning with the easycare cleaning aid is recommended if the nebulisation
time for the same fi ll volume of a given substance signifi cantly increases.
Additional information can be found in the current version of the corre-
sponding eFlow
®
word "prolonged nebulisation times" (www.pari.de/Products/Pari Inhalation
devices/eFlow).
Cleaning Aid für eFlow
Aerosol Head
appliance.
®
Instructions for Use in the Chapter "Troubleshooting", key
aerosol head with a built-in fun-
®
®
en
®
®
9