4 4
Configure the volume unit
1 Select Calibrate from the main menu .
2 Select Units Config from the "Diagnostics" submenu .
3 Select Volume Unit from the "Unit Config" submenu .
4 Configure the Volume Unit as required (liter is used in this
example) and then press Next .
L S T 3 0 0 | C O M PA C T U LT R A S O N I C L E V E L T R A N S M I T T E R | L S T 3 0 0 – E N R E V. E
Configure the linearization table
1 Select Calibrate from the main menu .
2 Select Linearization from the "Calibrate" submenu .
3 Set Max Volume from the "Linearization" submenu .
In this example, 1832 liters is set as the maximum volume (unit
already set in section "Configuring the volume unit") .
4 Enter the Linearization Table from the "Linearization"
submenu .