Use command ? to find out the settings of the active transmitter:
When the necessary settings have been given, close the line to
transmitters (the command closes all open lines):
CLOSE command is always given without address. If no lines are
open, there will be no response to the CLOSE command.

POLL mode

If a transmitter has been set to POLL mode, it will respond only to
commands which include its address:
Addresses from 1 to 99 can be used. According to the RS 485/422
standard a maximum of 32 devices can be connected on same bus, but
the number can be increased if the line length and/or baud rate is
reduced.
The line terminations must be dynamic; e.g. an RC circuit is used
instead of a simple resistor termination. Each RS 485 module has a
dynamic line termination so it can be used at the end of a line.
V
________________________________________________________________________ 113
AISALA
? <cr>
DMP248 / 1.01
CPU serial nr :
0
Keyboard type :
0
Address
:
7
Output units
:
metric
Baud P D S
:
4800 E 7 1 FDX
Serial mode
:
STOP
Output intrv. :
0 min
Mtim
:
32
Pressure :
1013.25
Analog outputs
Ch1
0.00 ...
10.00 V
Ch2
0.00 ...
10.00 V
Ch1
( RH )
lo
0.000 %RH
Ch1
( RH )
hi
100.000 %RH
Ch2
( T
)
lo
-20.000 'C
Ch2
( T
)
hi
180.000 'C
Transducer
:
PRB serial nr :
0
Calibr. date
:
0
>close
line closed
send 22
Td= -47.4 'C T= 29.1 'C