Serial Line Settings; Set Transmitter Address For Use In Poll Mode - Vaisala HUMICAP MMT310 User Manual

Moisture and temperature transmitter
Hide thumbs Also See for HUMICAP MMT310:
Table of Contents

Advertisement

User's Guide _______________________________________________________________________
40 ___________________________________________________________________ M210474EN-D
Example:
>smode run
Output mode
: RUN
>r
02:08:01.03 aw=
02:08:04.21 aw=
<ESC>
>smode stop
Output mode
: STOP

Serial Line Settings

Syntax: SERI b p d s<cr>
Where
b = bauds (300, 600, 1200, 2400, 4800, 9600,19200)
p = parity (n = none, e = even, o = odd)
d = data bits (7 or 8)
s = stop bits (1 or 2)
The settings can be changed one parameter at a time or all parameters at
once.
Example:
>seri o
4800 o 7 1
>seri 600 n 8 1
600 n 8 1
Set Transmitter Address for Use in POLL
Mode
Syntax: ADDR aa<cr>
Where
aa
=
Address (0 ... 99)
Example:
>addr
Address
: 0
>addr 1
Address
: 1
0.2 aw T=
-47.13 'C ...
0.2 aw T=
-47.16 'C ...
changing parity only
changing all parameters

Advertisement

Table of Contents
loading

Table of Contents