Close Serial Line In Poll Mode Command (Close); Set Echoing Mode Command (Echo) - Vaisala Spectracap OMT355 User Manual

Oxygen transmitter
Table of Contents

Advertisement

User's Guide ______________________________________________________________________
70 __________________________________________________________________ M210735EN-D
Close Serial Line in Poll Mode
Command (CLOSE)
This command closes a device and switches it into POLL mode. Unless
an addressable command is issued all output is suppressed until a reset
occurs or the OPEN command is used. If serial mode is set to POLL
using SMODE command and the setting is saved to EEPROM (SAVE
command), the device wakes up after a reset still in POLL mode and
output is suppressed also on start-up.
Syntax: CLOSE<cr>
Example:
>close
line closed
Set Echoing Mode Command
(ECHO)
This command sets the echoing mode. In RS232C mode the device
echoes everything back to the user by default. In RS-485 mode echoing
is automatically disabled. In the example below, the two commands in
italics are typed by the user but not seen on the screen unless using local
echo.
Syntax: ECHO [on/off]<cr>
where
on
= Echo on
off
= Echo off

Hide quick links:

Advertisement

Table of Contents
loading