4.
CHECKING THE SERIAL PORT NETWORK OPERATION
teitsurement readings are normally asked with the transmitter in POLL
mode; the command must include the address of the transmitter. If the settings
of the transmitter need to be changed, the transmitter is switched to STOP
mode with command OPEN; commands can then be given without address.
When the line to the transmitter is closed, it returns to POLL mode.
STOP mode
Open the line to the transmitter:
opesn 22<ccz.
114,
22 line,
openred lo
operator eoarnands
Transmitter no. 22 is now temporarily set to STOP mode; it accepts com-
mands sent without address until CLOSE command is given. Individual set-
tings can now be easily modified. Do not open more than one line at a time.
Use command ? to find out the settings of the active transmitter:
7cr
IM4' 240 / 1.02
CPU serial
nr
0
I:eyboacd type
:
0B
Addvoss
.
7
0u" tut
uniLs
I
suetric
Baud P D S
:
4800 Z 7 I1FPX
M.srial moid.
STOP
nitpur iorrv.
.
0 min
1, i.
32
Pressure
1013.25
An alog outputs
Chl
0.00
...
10.00 V
Ch2
0.00
...
10.00 V
Chl
I R1 1
Io
0.000
tRIn
Chl
A 1ll )
hi
100.000 Intl
Ch2
( T
I
to
-20.000
'C
Cli?
T
hi
190.000 -C
Transducer
PRI serial
nr ,
0
Cr:lib,".
date
a
0
When the necessary settings have been given, close the line to transmitters
(the command closes all open lines):
>croso
line closed
CLOSE command is always given without address. If no line is open, there
will be no response to the CLOSE commands.
POLL mode
If a transmitter has been set to POLL mode. it will respond only to commands
sent with its address:
send 22
OH.
24.41
111 T.
29.1
"
A
93
X,
.4