Pro. This USB 2.0 cord should only be unplugged
from the Q-View Imager Pro once Q-View Software
is closed and the Q-View Imager Pro is powered off.

Power Switch

The power switch is located on the back of the device
above the power plug receptacle and the fuse box.
This switch controls the power to the Q-View Imager
Pro and NOT the PC. To turn on the Q-View Imager
Pro, push the power switch to the "|" position. To turn
off the Q-View Imager Pro, push the power switch to
the "O" position.

Fuse

The fuse is located on the back of the Q-View Imager
Pro in-between the power plug receptacle and the
power switch. The fuse is 250V/10A/T. If the device
is not powering up when the switch is turned on, the
user can check the fuse.
Carefully remove the fuse receptacle. Inside the
receptacle, there are two fuses; one is inline and the
other is a backup. If the main fuse, which is connected
to the circuitry, is broken, blown, or discolored, the
user can switch the fuses. If you run out of fuses,
please contact Quansys technical support.

ROUTINE MAINTENANCE

Cleaning

Cleaning is recommended before every use. Ensure
that the glass, glass cover, and tray are free from
dust and any chemicals. To clean the device, use 70
percent ethanol or warm water. Do not use scouring
cleaners or abrasive pads as this can scratch the
glass. Be careful not to spill any cleaning materials
inside the Q-View Imager Pro.
Regularly inspect all cables for fraying or loose
connections.