RTLC20/FASTRACK carrier installation
Slide RTLC20 scale into the FASTRACK carrier
6
ensuring the scale is fed under the projections
as shown.
Scale can be installed manually by either
pulling or pushing it through the
FASTRACK carrier.
Alternatively use the optional
scale installation tool
(A-9589-0420) as shown, for
easy installation.
7
Optional: fix self‑adhesive end covers
ensuring a gap of at least 1 mm between
the end of the scale and the end cover.
Ensure a gap > 1 mm
8
Clean FASTRACK and scale
using a lint-free cloth.
VIONiC RTLC20/FASTRACK installation guide
(Continued)
NOTE: For instructions on how to use the scale
installation tool, download the RTL scale installation
tool (A-9589-0420) user guide (Renishaw
part no. M-9589-9101) from the website at
www.renishaw.com/encoderinstallationguides
Projections
IMPORTANT: If manually installing
the scale using fingers, suitable
gloves should be worn to protect
against injury from sharp edges.
Scale datum
The datum clamp fixes the RTLC20 scale rigidly to the substrate at the location chosen.
CAUTION: The metrology of the system may be compromised if the datum clamp is not used.
The datum clamp does not need to be fitted adjacent to a reference mark. It can be positioned anywhere
along the axis depending upon the customer's requirements.
Mechanical clamp
Reference mark
selector magnet
15 ±1
IN-TRAC reference mark
25
Bolted datum clamp
(A-9589-0077)
3
12
M3 × 6 low head screw,
supplied with clamp
Additional screws available (pack of 25 A-9584-2047).
Adhesive clamp
Using dispensing tip P-TL50-0209 apply Loctite 435
between the scale and the FASTRACK carrier so it
wicks underneath adjacent to user selected datum
location as shown.
NOTE: Drawing shows the scale datum adjacent to
the chosen reference mark.
NOTE: Only apply Loctite 435 in
these gaps to ensure best bond.
Loctite 435 will wick under the
scale to lock it to the substrate.
2.3
1.7
5 min
thread
depth
Do not place clamp tip
NOTE: When the reference mark and
over shaded area
datum clamp are aligned as shown,
the reference output will be positionally
repeatable with respect to the substrate.
12
16
6