Data Port Interface Protocol

Set Date and Time (Option 2)

Dump Trend (Option 3)

Dump Error Log (Option 4)

Exit Interactive Mode (Option 5)

138
When the instrument is shipped from the factory, the date and time are set to the
local time zone during manufacturing. If the battery has been removed or
disconnected, the time clock will not reflect the actual date and time. After battery
power has been restored, reset the time clock.
Option 2 is used to change the date and time via the computer. The format for date
and time is DD-MM-YY HH:MM:SS. Move the cursor under the value to be
changed and enter the new value.
Option 3 outputs current trend information. Up to 48 hours of trend information
can be viewed. Information presented includes:
Instrument Type
Software Revision Level
Printout Type
Alarm Limits
Date And Time
%SpO
2
Pulse Rate
Pulse Amplitude
Option 4 outputs a list of all of the in memory. The information that can be
viewed includes instrument type, software revision level, printout type, time of
printout, operating time of the recorded error, error number, task number, address,
and count. This option is intended for Nellcor's field service personnel.
Option 5 closes the interactive mode and returns the data port to normal operation.