Download Print this page

Hach 5500sc Installation Manual page 3

Ammonia/monochloramine
Hide thumbs Also See for 5500sc:

Advertisement

Available languages

Available languages

  • ENGLISH, page 3
Table of contents
Specifications
on page 3
General information
on page 4
Installation
on page 8
Specifications
Specifications are subject to change without notice.
Specification
Dimensions (W x D x H)
Enclosure
Weight
Mounting
Protection class
Pollution degree/installation
category
Power requirements
Operating temperature
Operating humidity
Storage temperature
4–20 mA outputs
Alarm relay outputs
Digital inputs
Fuses
Fittings
Table 1 General specifications
Details
452 x 360 x 804 mm (17.8 x 14.2 x 31.7 in.)
Rating: NEMA 4x/IP56
Material: PC/ABS case, PC door, PC hinges and latches, 316 SST hardware
Indoor use only. Keep away from direct sunlight.
20.5 kg (45.2 lb) without reagents and standards, 30 kg (66.1 lb) with reagents,
standards and cleaning solution
Wall, panel or table
I
2/II
AC: 100–240 VAC, 50/60 Hz
Instrument: 0.5 A nominal, 8.3 A maximum; accessory output 100–240 Vac,
5.0 A maximum
Connection: 1.0 to 1.3 mm
recommended; field wiring insulation must be rated to 65°C (149 ºF) minimum
5 to 45 °C (41 to 113 °F)
5 to 95% non-condensing
–20 to 60 °C (–4 to 140 °F)
Four; load impedance: 600 Ω maximum
Connection: 0.644 to 1.29 mm
(22 to 20 AWG) recommended, twisted pair shielded wire
Four; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC
maximum
Connection: 1.0 to 1.29 mm
stranded recommended
Four; connection: 0.644 to 1.29 mm
(22 to 20 AWG) stranded (isolated DC voltage input or an open-collector/relay
contact closure input) recommended
Input power—AC: T 1.6 A, 250 VAC; DC: T 6.3 A, 250 VAC
Output power—AC: T 5.0 A, 250 VAC; DC: T 1.6 A, 250 VAC
Alarm relay outputs: T 5.0 A, 250 V
Sample line and sample bypass drain: 6 mm OD push-to-connect fitting for
plastic tubing
Chemical and case drains: 11 mm (7/16 in.) ID slip-on fitting for soft plastic
tubing
2
(18 to 16 AWG) wire, 1.0 mm
2
(22 to 16 AWG) wire, 0.644 to 0.812 mm
2
(18 to 16 AWG) wire, 1.0 mm
2
(22 to 16 AWG) wire, 0.644 to 0.812 mm
2
(18 AWG) stranded
2
2
(18 AWG)
2
English 3

Advertisement

loading